{"id":2541,"date":"2020-05-04T18:58:03","date_gmt":"2020-05-04T18:58:03","guid":{"rendered":"https:\/\/liquidinstruments.com\/?p=2541"},"modified":"2024-10-18T18:53:47","modified_gmt":"2024-10-18T18:53:47","slug":"mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact","status":"publish","type":"post","link":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/","title":{"rendered":"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger","gt_translate_keys":[{"key":"rendered","format":"text"}]},"content":{"rendered":"<p><em>Updated April 24, 2023<\/em><\/p>\n<p>In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain gauge. We are aiming to drop a steel ball onto an aluminum plate, measure the strain of this roughly 100 ms event, and observe any resulting mechanical oscillations. In this example, we use Moku:Lab, but all Moku devices are capable of performing these measurements.<\/p>\n<div style=\"text-align: center; padding-bottom: 40px;\">\n\n<\/div>\n<hr style=\"margin: 50px 0 50px 0 !important;\" \/>\n<h2>Moku:Lab Data Logger instrument<\/h2>\n<p>The Moku:Lab Data Logger instrument records time series voltages from 1 or 2 channels at rates from 1 sample per second up to 1 MSa\/s. The data can be logged to RAM or a removeable SD card in a variety of formats including MATLAB\u2019s .MAT file format. With sufficient storage, very lengthy logs of up to 240 hours can be created. The resulting logs can be shared to email or cloud services such as Dropbox. The Moku:Lab Data Logger also includes an embedded Waveform Generator, useful for providing triggers or stimulus. Complete details on the Data Logger are available in the user manual [1].<\/p>\n<h3>Experimental goal<\/h3>\n<p>We are going to drop a steel ball onto an aluminum plate and observe and record the strain of the impact and any resulting mechanical oscillations in the plate. We will use a strain gauge and strain gauge amplifier in conjunction with the Moku:Lab Oscilloscope and Data Logger instruments.<\/p>\n<h3>What is strain?<\/h3>\n<p>In mechanical test and measurement, strain is defined as the change in length of an object relative to its unstrained length. Strain (\u03b5) is defined thus:<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2558\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/01-Formula-01.jpg\" alt=\"Strain (\u03b5) definition \" width=\"500\" height=\"150\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/01-Formula-01.jpg 1000w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/01-Formula-01-300x90.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/01-Formula-01-768x230.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/01-Formula-01-600x180.jpg 600w\" sizes=\"(max-width: 500px) 100vw, 500px\" \/><\/p>\n<p>When an object is subject to compressive forces, its length reduces, while the other two dimensions will slightly increase. Equally, when an object is subject to tensile force, its length increases, while the other two dimensions will reduce (Figure 1).<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2559\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-scaled.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-02-Figure-01\" width=\"599\" height=\"393\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-scaled.jpg 2560w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-300x197.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-1024x672.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-768x504.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-1536x1008.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-2048x1344.jpg 2048w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/02-Figure-01-600x394.jpg 600w\" sizes=\"(max-width: 599px) 100vw, 599px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>1<\/strong><strong>: <\/strong>Effect of tension and compression<\/em><\/p>\n<p>When an object is bent, such as during an impact from above, the top surface will be in compression while the bottom surface is in tension (Figure 2).<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2560\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-03-Figure-02\" width=\"599\" height=\"337\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02.jpg 1920w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02-300x169.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02-1024x576.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02-768x432.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02-1536x864.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/03-Figure-02-600x338.jpg 600w\" sizes=\"(max-width: 599px) 100vw, 599px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>2<\/strong><strong>: <\/strong>Effect of bending force<\/em><\/p>\n<p>In practical situations, measured strain is very small and hence the unit-less measurement millistrain (10<sup>-3<\/sup>) or microstrain (10<sup>-6<\/sup>) is often used.<\/p>\n<h3>Strain gauge background<\/h3>\n<p>A strain gauge is a mechanical-electrical sensor that allows us to measure mechanical strain. It relies upon the physical properties of conductance of an electrical conductor. The electrical resistance of a conductor of a given material is dependent upon the material\u2019s conductivity and the conductor length and thickness. Resistance is defined by:<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2561\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/04-Formula-02.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-04-Figure-02\" width=\"500\" height=\"150\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/04-Formula-02.jpg 1000w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/04-Formula-02-300x90.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/04-Formula-02-768x230.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/04-Formula-02-600x180.jpg 600w\" sizes=\"(max-width: 500px) 100vw, 500px\" \/><\/p>\n<p>w here R = resistance (\u03a9), l = length (m) and A (cross sectional area m<sup>2<\/sup>). \u03c1\u00a0is the electrical resistivity of the material (\u03a9-m).<\/p>\n<p>As a conductor is stretched, it will lengthen and narrow, both increasing the electrical resistance. Equally, as a conductor is compressed, it will shorten and thicken, both decreasing the electrical resistance. Thus, mechanical strain can be directly translated into electrical resistance.<\/p>\n<p>Resistivity (\u03c1) of copper is 1.72 e<sup>-8<\/sup> \u03a9-m; for a fine wire of diameter 0.1 mm in a compact size, of say, length 1 cm, the resistance is 22 m\u03a9. If we are aiming to measure millistrain, the measurement challenge is one of measuring very small resistance changes in the order of \u00b5\u03a9. To make the resistance measurement simpler, a longer conductor length and ideally a material of higher resistivity is desirable.<\/p>\n<p>A strain gauge implements this fine wire in a small, compact area by using a foil zig-zag pattern in order to maximize conductor length in a small footprint.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2562\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03.jpeg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-05-Figure-03\" width=\"557\" height=\"304\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03.jpeg 2344w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03-300x163.jpeg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03-1024x558.jpeg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03-768x418.jpeg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03-1536x837.jpeg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03-2048x1116.jpeg 2048w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/05-Figure-03-600x327.jpeg 600w\" sizes=\"(max-width: 557px) 100vw, 557px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>3<\/strong><strong>: <\/strong>Photograph of a strain gauge, 4mm x 6mm<\/em><\/p>\n<p>In order to measure the resistance of the strain gauge, it is necessary to pass a current through the wire; this will cause some self-heating to occur. The temperature coefficient of copper, at room temperature, is 4.29 e<sup>-3<\/sup> ppm\/\u00b0C. So, for a modest increase of 10\u00b0C, the resistance changes 4.29 %. This negatively impacts the ability to measure strain accurately.<\/p>\n<p>Commercially available strain gauges typically use the alloy constantan, a copper-nickel alloy.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2563\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/06-Table-01.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-06-Table-01\" width=\"557\" height=\"346\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/06-Table-01.jpg 1041w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/06-Table-01-300x186.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/06-Table-01-1024x636.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/06-Table-01-768x477.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/06-Table-01-600x373.jpg 600w\" sizes=\"(max-width: 557px) 100vw, 557px\" \/><\/p>\n<p>Constantan has a resistivity (\u03c1) of 4.9e<sup>-7<\/sup> (28 times higher than copper) and notably a temperature coefficient of 8 e<sup>-6<\/sup> ppm\/\u00b0C, &gt;500 times more stable over temperature. A 10\u00b0C rise results in just 0.008 % change in resistance, a very useful benefit compared to copper.<\/p>\n<p>Still, a small gauge made of a fine constantan zig-zag would have a typical unstrained resistance of just 100 \u03a9 or so; and to measure millistrain means we need to measure fractions of 1 \u03a9. Typical strain gauges have unstrained resistance of 120 \u03a9 or 350 \u03a9.<\/p>\n<h3>Wheatstone bridge<\/h3>\n<p>Small resistance changes are typically made with a Wheatstone bridge circuit. Such a circuit consists of two voltage dividers in parallel driven by an excitation voltage. A high impedance voltmeter measures the potential difference between the two voltage dividers. This effectively measures the resistance difference across the two arms and can result in very precise measurements of resistance changes. Further, the measurement does not vary due to voltage variations in the excitation source and provided the resistors are matched and of low temperature coefficients, this arrangement greatly reduces temperature effects.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2564\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/07-Figure-04.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-07-Figure-04\" width=\"501\" height=\"451\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/07-Figure-04.jpg 1141w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/07-Figure-04-300x270.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/07-Figure-04-1024x921.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/07-Figure-04-768x691.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/07-Figure-04-600x540.jpg 600w\" sizes=\"(max-width: 501px) 100vw, 501px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>4<\/strong><strong>: <\/strong>Wheatstone bridge<\/em><\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2548\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/08-Figure-05.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-08-Figure-05\" width=\"499\" height=\"472\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/08-Figure-05.jpg 1064w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/08-Figure-05-300x284.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/08-Figure-05-1024x968.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/08-Figure-05-768x726.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/08-Figure-05-600x567.jpg 600w\" sizes=\"(max-width: 499px) 100vw, 499px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>5<\/strong><strong>: <\/strong>Wheatstone bridge with strain gauge<\/em><\/p>\n<p>Assuming the bridge is in balance, that is, there is 0 volts at the voltmeter (V), we can say:<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2549\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/09-Formula-03.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-09-Formula-03\" width=\"500\" height=\"150\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/09-Formula-03.jpg 1000w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/09-Formula-03-300x90.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/09-Formula-03-768x230.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/09-Formula-03-600x180.jpg 600w\" sizes=\"(max-width: 500px) 100vw, 500px\" \/><\/p>\n<p>Assuming that the voltmeter is high impedance and negligible current flows through it, we can replace R4 with a strain gauge and state:<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2550\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/10-Formula-04.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-10-Formula-04\" width=\"500\" height=\"150\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/10-Formula-04.jpg 1000w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/10-Formula-04-300x90.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/10-Formula-04-768x230.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/10-Formula-04-600x180.jpg 600w\" sizes=\"(max-width: 500px) 100vw, 500px\" \/><\/p>\n<p>Where Ve is the source excitation voltage and Rx the resistance of the strain gauge. Rearranging for Rx (see [3] for fuller discussion) gives:<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2551\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/11-Formula-05.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-11-Formula-05\" width=\"500\" height=\"150\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/11-Formula-05.jpg 1000w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/11-Formula-05-300x90.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/11-Formula-05-768x230.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/11-Formula-05-600x180.jpg 600w\" sizes=\"(max-width: 500px) 100vw, 500px\" \/><\/p>\n<p>Still, in order to measure millistrain we will be measuring millivolts and so we need a suitable amplifier. We used a readily available dedicated strain gauge amplifier [2] and configured it to 0.06 mV\/V (equal to a gain of 16,667) to give a suitable output voltage range for the Moku:Lab inputs (Figure 6).<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2552\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-12-Figure-06\" width=\"507\" height=\"286\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06.jpg 1920w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06-300x169.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06-1024x576.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06-768x432.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06-1536x864.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/12-Figure-06-600x338.jpg 600w\" sizes=\"(max-width: 507px) 100vw, 507px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>6<\/strong><strong>: <\/strong>Strain gauge, amplifier and Moku:Lab<\/em><\/p>\n<p>&nbsp;<\/p>\n<hr \/>\n<h2>Experimental setup<\/h2>\n<p>The experimental setup is shown in Figure 7. An aluminum plate is struck by a steel ball dropped from a height of 1 m. As the ball impacts the plate, we would expect to see an initial deflection of the bar (hence strain) followed by a brief, millisecond series of oscillation of the plate.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2553\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-13-Figure-07\" width=\"600\" height=\"337\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07.jpg 1920w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07-300x169.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07-1024x576.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07-768x432.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07-1536x864.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/13-Figure-07-600x338.jpg 600w\" sizes=\"(max-width: 600px) 100vw, 600px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>7<\/strong><strong>: <\/strong>Experimental setup<\/em><\/p>\n<h3>Moku:Lab Data Logger setup<\/h3>\n<p>The Moku:Lab Data Logger was used to capture the output of the strain gauge during the ball drop. The data logger was set to high speed (1 MSa\/s) and the impact captured. The log file was shared directly from Moku:Lab via ethernet to a network connected PC as a \u201c.MAT\u201d format and then imported into MATLAB for viewing.<\/p>\n<p>&nbsp;<\/p>\n<hr \/>\n<h2>Result and observations<\/h2>\n<p>Figure 8 shows the Moku:Lab time series plot of the ball impact as captured on the Moku:Lab Oscilloscope. This trace extends for approximately 11 ms and shows a large initial impact strain of around 1.5 ms followed by a series of oscillations of period of approximately 1.2 ms (or 833 Hz). Visually, there appears to be some higher frequency components.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2554\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/14-Figure-08.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-14-Figure-08\" width=\"600\" height=\"462\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/14-Figure-08.jpg 1360w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/14-Figure-08-300x231.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/14-Figure-08-1024x789.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/14-Figure-08-768x592.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/14-Figure-08-600x462.jpg 600w\" sizes=\"(max-width: 600px) 100vw, 600px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>8<\/strong><strong>: <\/strong>Oscilloscope plot of impact<\/em><\/p>\n<p>The .MAT file produced from the Moku:Lab Data Logger was then imported into MATLAB and the longer time series is show in Figure 9.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2555\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-scaled.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-15-Figure-09\" width=\"611\" height=\"306\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-scaled.jpg 2560w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-300x150.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-1024x513.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-768x385.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-1536x769.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-2048x1026.jpg 2048w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/15-Figure-09-600x300.jpg 600w\" sizes=\"(max-width: 611px) 100vw, 611px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>9<\/strong><strong>:\u00a0<\/strong>Impact and oscillations from the Data Logger<\/em><\/p>\n<p>This shows the large strain of the initial impact (3 V peak deviation), followed by more than 3.5 seconds of decaying oscillations. The abrupt amplitude changes at around 3.7 and 3.8 seconds correspond to the plate bouncing up and hitting the floor twice. Figure 10 shows the response from 3.45 s to 3.455 s and the oscillations previously noted on the Oscilloscope.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2556\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-scaled.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-10-Figure-10\" width=\"624\" height=\"265\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-scaled.jpg 2560w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-300x127.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-1024x435.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-768x326.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-1536x653.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-2048x870.jpg 2048w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/16-Figure-10-600x255.jpg 600w\" sizes=\"(max-width: 624px) 100vw, 624px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>10<\/strong><strong>: <\/strong>Close up of oscillation from the Data Logger<\/em><\/p>\n<p>Since Moku:Lab allowed us to easily import the higher resolution time series in to MATLAB, we used MATLAB to produce a frequency plot (FFT) of the datalog. This is shown in Figure 11 with significant peaks of energy at 950 Hz and 1700 Hz.<\/p>\n<p style=\"text-align: center;\"><img decoding=\"async\" class=\"alignnone wp-image-2557\" src=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-scaled.jpg\" alt=\"20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-17-Figure-11\" width=\"618\" height=\"262\" srcset=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-scaled.jpg 2560w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-300x127.jpg 300w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-1024x435.jpg 1024w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-768x326.jpg 768w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-1536x652.jpg 1536w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-2048x870.jpg 2048w, https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/04\/17-Figure-11-600x255.jpg 600w\" sizes=\"(max-width: 618px) 100vw, 618px\" \/><\/p>\n<p style=\"padding-left: 15%; padding-right: 15%; text-align: center; font-size: 18px;\"><em><strong>Figure<\/strong> <strong>11<\/strong><strong>: <\/strong>FFT of impact datalog<\/em><\/p>\n<p>The 950 Hz peak corresponds with the earlier estimate of 800 Hz from the Moku:Lab Oscilloscope (Figure 8). There is very little oscillation energy above 1700 Hz.<\/p>\n<p>&nbsp;<\/p>\n<hr \/>\n<h2><\/h2>\n<h2>Summary<\/h2>\n<p>In this application note we have explained the basic concepts of mechanical strain and a common measurement method. We then used the Moku:Lab Data Logger to record the strain of a mechanical impact of a steel ball dropped onto an aluminum plate.<\/p>\n<p>Moku:Lab allowed for easy data sharing via Dropbox of a data log in MATLAB .MAT file format and the subsequent viewing of the time series and plotting of the frequency components.<\/p>\n<p>Moku:Lab has MATLAB APIs and a future development of this experiment could be to automate the data collection and analysis entirely in MATLAB using Moku:Lab\u2019s network connection.<\/p>\n<p>&nbsp;<\/p>\n<hr style=\"margin: 50px 0 50px 0 !important;\" \/>\n<h2>References<\/h2>\n<p>[1] Moku:Lab Data Logger User Manual : <a href=\"http:\/\/download.liquidinstruments.com\/documentation\/manual\/instrument\/mokulab\/User%20Manual%20-%20MokuLab%20Data%20Logger.pdf\">http:\/\/download.liquidinstruments.com\/documentation\/manual\/instrument\/mokulab\/User%20Manual%20%20MokuLab%20Data%20Logger.pdf <\/a><\/p>\n<p>[2] SGA-A strain gauge amplifier with integrated Wheatstone bridge : <a href=\"https:\/\/www.mantracourt.com\/products\/analog-output-signal-conditioners\/strain-gauge-amplifier\">https:\/\/www.mantracourt.com\/products\/analog-output-signal-conditioners\/strain-gauge-amplifier<\/a><\/p>\n<p>[3] Wikipedia description and mathematical discuss of Wheatstone bridge <a href=\"https:\/\/en.wikipedia.org\/wiki\/Wheatstone_bridge\">https:\/\/en.wikipedia.org\/wiki\/Wheatstone_bridge <\/a><\/p>\n<hr style=\"margin: 50px 0 50px 0 !important;\" \/>\n<h2>Questions or comments?<\/h2>\n<p>Please contact us at <a href=\"mailto:support@liquidinstruments.com\">support@liquidinstruments.com<\/a><\/p>\n","protected":false,"gt_translate_keys":[{"key":"rendered","format":"html"}]},"excerpt":{"rendered":"<p>Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain gauge. We are aiming to drop a steel ball onto an aluminum plate, measure the strain of this roughly 100 ms event, and observe any resulting mechanical oscillations. In [&hellip;]<\/p>\n","protected":false,"gt_translate_keys":[{"key":"rendered","format":"html"}]},"author":14,"featured_media":3920,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"content-type":"","footnotes":""},"categories":[5],"tags":[],"class_list":["post-2541","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-application-notes","site-category-data-logger","site-category-mokulab"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.0 (Yoast SEO v27.0) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>Capturing high-speed mechanical impact using the Moku:Lab Data Logger<\/title>\n<meta name=\"description\" content=\"Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger\" \/>\n<meta property=\"og:description\" content=\"Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain\" \/>\n<meta property=\"og:url\" content=\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\" \/>\n<meta property=\"og:site_name\" content=\"Liquid Instruments\" \/>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/LiquidInstruments\/\" \/>\n<meta property=\"article:published_time\" content=\"2020-05-04T18:58:03+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2024-10-18T18:53:47+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg\" \/>\n\t<meta property=\"og:image:width\" content=\"1080\" \/>\n\t<meta property=\"og:image:height\" content=\"720\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/jpeg\" \/>\n<meta name=\"author\" content=\"joel\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:creator\" content=\"@liquidinstrmnts\" \/>\n<meta name=\"twitter:site\" content=\"@liquidinstrmnts\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"joel\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"11 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\/\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#article\",\"isPartOf\":{\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\"},\"author\":{\"name\":\"joel\",\"@id\":\"https:\/\/liquidinstruments.com\/#\/schema\/person\/603af464a65f123df56503f020b73dba\"},\"headline\":\"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger\",\"datePublished\":\"2020-05-04T18:58:03+00:00\",\"dateModified\":\"2024-10-18T18:53:47+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\"},\"wordCount\":1486,\"commentCount\":0,\"publisher\":{\"@id\":\"https:\/\/liquidinstruments.com\/#organization\"},\"image\":{\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg\",\"articleSection\":[\"Application notes\"],\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\",\"url\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\",\"name\":\"Capturing high-speed mechanical impact using the Moku:Lab Data Logger\",\"isPartOf\":{\"@id\":\"https:\/\/liquidinstruments.com\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage\"},\"image\":{\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage\"},\"thumbnailUrl\":\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg\",\"datePublished\":\"2020-05-04T18:58:03+00:00\",\"dateModified\":\"2024-10-18T18:53:47+00:00\",\"description\":\"Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain\",\"breadcrumb\":{\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage\",\"url\":\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg\",\"contentUrl\":\"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg\",\"width\":1080,\"height\":720,\"caption\":\"Oscilloscope plot of impact\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\/\/liquidinstruments.com\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\/\/liquidinstruments.com\/#website\",\"url\":\"https:\/\/liquidinstruments.com\/\",\"name\":\"Liquid Instruments\",\"description\":\"\",\"publisher\":{\"@id\":\"https:\/\/liquidinstruments.com\/#organization\"},\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\/\/liquidinstruments.com\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Organization\",\"@id\":\"https:\/\/liquidinstruments.com\/#organization\",\"name\":\"Liquid Instruments\",\"url\":\"https:\/\/liquidinstruments.com\/\",\"logo\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\/\/liquidinstruments.com\/#\/schema\/logo\/image\/\",\"url\":\"https:\/\/i0.wp.com\/liquidinstruments.com\/wp-content\/uploads\/2020\/10\/BrandMark-Preferred-RGB-Color.png?fit=1000%2C924&ssl=1\",\"contentUrl\":\"https:\/\/i0.wp.com\/liquidinstruments.com\/wp-content\/uploads\/2020\/10\/BrandMark-Preferred-RGB-Color.png?fit=1000%2C924&ssl=1\",\"width\":1000,\"height\":924,\"caption\":\"Liquid Instruments\"},\"image\":{\"@id\":\"https:\/\/liquidinstruments.com\/#\/schema\/logo\/image\/\"},\"sameAs\":[\"https:\/\/www.facebook.com\/LiquidInstruments\/\",\"https:\/\/x.com\/liquidinstrmnts\",\"https:\/\/www.instagram.com\/liquidinstruments\/\",\"https:\/\/www.linkedin.com\/company\/liquidinstruments\/\",\"https:\/\/www.youtube.com\/c\/LiquidInstruments\",\"https:\/\/vimeo.com\/liquidinstruments\"],\"hasMerchantReturnPolicy\":{\"@type\":\"MerchantReturnPolicy\",\"merchantReturnLink\":\"https:\/\/liquidinstruments.com\/support\/warranty-repairs-and-service\/\"}},{\"@type\":\"Person\",\"@id\":\"https:\/\/liquidinstruments.com\/#\/schema\/person\/603af464a65f123df56503f020b73dba\",\"name\":\"joel\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\/\/liquidinstruments.com\/#\/schema\/person\/image\/\",\"url\":\"https:\/\/secure.gravatar.com\/avatar\/61d26ea66bf9fb0a2379e1e6e8febb312981e656713bff920c55016dfe82495d?s=96&d=wavatar&r=g\",\"contentUrl\":\"https:\/\/secure.gravatar.com\/avatar\/61d26ea66bf9fb0a2379e1e6e8febb312981e656713bff920c55016dfe82495d?s=96&d=wavatar&r=g\",\"caption\":\"joel\"}}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"Capturing high-speed mechanical impact using the Moku:Lab Data Logger","description":"Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/","og_locale":"en_US","og_type":"article","og_title":"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger","og_description":"Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain","og_url":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/","og_site_name":"Liquid Instruments","article_publisher":"https:\/\/www.facebook.com\/LiquidInstruments\/","article_published_time":"2020-05-04T18:58:03+00:00","article_modified_time":"2024-10-18T18:53:47+00:00","og_image":[{"width":1080,"height":720,"url":"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg","type":"image\/jpeg"}],"author":"joel","twitter_card":"summary_large_image","twitter_creator":"@liquidinstrmnts","twitter_site":"@liquidinstrmnts","twitter_misc":{"Written by":"joel","Est. reading time":"11 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#article","isPartOf":{"@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/"},"author":{"name":"joel","@id":"https:\/\/liquidinstruments.com\/#\/schema\/person\/603af464a65f123df56503f020b73dba"},"headline":"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger","datePublished":"2020-05-04T18:58:03+00:00","dateModified":"2024-10-18T18:53:47+00:00","mainEntityOfPage":{"@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/"},"wordCount":1486,"commentCount":0,"publisher":{"@id":"https:\/\/liquidinstruments.com\/#organization"},"image":{"@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage"},"thumbnailUrl":"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg","articleSection":["Application notes"],"inLanguage":"en-US","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#respond"]}]},{"@type":"WebPage","@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/","url":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/","name":"Capturing high-speed mechanical impact using the Moku:Lab Data Logger","isPartOf":{"@id":"https:\/\/liquidinstruments.com\/#website"},"primaryImageOfPage":{"@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage"},"image":{"@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage"},"thumbnailUrl":"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg","datePublished":"2020-05-04T18:58:03+00:00","dateModified":"2024-10-18T18:53:47+00:00","description":"Updated April 24, 2023 In this application note we describe the use of the Moku Data Logger to capture a high-speed mechanical impact using a strain","breadcrumb":{"@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#primaryimage","url":"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg","contentUrl":"https:\/\/liquidinstruments.com\/wp-content\/uploads\/2020\/05\/20-0429_AppNoteBlogPost-HighSpeed_StrainGauge_Datalogger-00-Hero-Outer.jpg","width":1080,"height":720,"caption":"Oscilloscope plot of impact"},{"@type":"BreadcrumbList","@id":"https:\/\/liquidinstruments.com\/application-notes\/mokulab-and-data-logging-using-the-data-logger-to-capture-high-speed-mechanical-impact\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/liquidinstruments.com\/"},{"@type":"ListItem","position":2,"name":"Capturing a high-speed mechanical impact using the Moku:Lab Data Logger"}]},{"@type":"WebSite","@id":"https:\/\/liquidinstruments.com\/#website","url":"https:\/\/liquidinstruments.com\/","name":"Liquid Instruments","description":"","publisher":{"@id":"https:\/\/liquidinstruments.com\/#organization"},"potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/liquidinstruments.com\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Organization","@id":"https:\/\/liquidinstruments.com\/#organization","name":"Liquid Instruments","url":"https:\/\/liquidinstruments.com\/","logo":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/liquidinstruments.com\/#\/schema\/logo\/image\/","url":"https:\/\/i0.wp.com\/liquidinstruments.com\/wp-content\/uploads\/2020\/10\/BrandMark-Preferred-RGB-Color.png?fit=1000%2C924&ssl=1","contentUrl":"https:\/\/i0.wp.com\/liquidinstruments.com\/wp-content\/uploads\/2020\/10\/BrandMark-Preferred-RGB-Color.png?fit=1000%2C924&ssl=1","width":1000,"height":924,"caption":"Liquid Instruments"},"image":{"@id":"https:\/\/liquidinstruments.com\/#\/schema\/logo\/image\/"},"sameAs":["https:\/\/www.facebook.com\/LiquidInstruments\/","https:\/\/x.com\/liquidinstrmnts","https:\/\/www.instagram.com\/liquidinstruments\/","https:\/\/www.linkedin.com\/company\/liquidinstruments\/","https:\/\/www.youtube.com\/c\/LiquidInstruments","https:\/\/vimeo.com\/liquidinstruments"],"hasMerchantReturnPolicy":{"@type":"MerchantReturnPolicy","merchantReturnLink":"https:\/\/liquidinstruments.com\/support\/warranty-repairs-and-service\/"}},{"@type":"Person","@id":"https:\/\/liquidinstruments.com\/#\/schema\/person\/603af464a65f123df56503f020b73dba","name":"joel","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/liquidinstruments.com\/#\/schema\/person\/image\/","url":"https:\/\/secure.gravatar.com\/avatar\/61d26ea66bf9fb0a2379e1e6e8febb312981e656713bff920c55016dfe82495d?s=96&d=wavatar&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/61d26ea66bf9fb0a2379e1e6e8febb312981e656713bff920c55016dfe82495d?s=96&d=wavatar&r=g","caption":"joel"}}]}},"gt_translate_keys":[{"key":"link","format":"url"}],"_links":{"self":[{"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/posts\/2541","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/users\/14"}],"replies":[{"embeddable":true,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/comments?post=2541"}],"version-history":[{"count":2,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/posts\/2541\/revisions"}],"predecessor-version":[{"id":20632,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/posts\/2541\/revisions\/20632"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/media\/3920"}],"wp:attachment":[{"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/media?parent=2541"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/categories?post=2541"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/liquidinstruments.com\/wp-json\/wp\/v2\/tags?post=2541"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}